Even well-calibrated qPCR cyclers can introduce unnoticed measurement errors. This short note outlines a practical, low-cost method for assessing instrument performance through run homogeneity testing - a functional validation approach that evaluates the optical and thermal consistency of your qPCR runs. Applicable to any laboratory setup, it enables regular trend analysis and quality assurance without specialized equipment.
The challenge of accurate qPCR data
There is the saying "quantitative PCR is easy to perform, but hard to do it right". With high quality instruments, robust reagents and pre-designed (pre-validated) assays, even novice users can easily generate qPCR data. The challenge is in ensuring that the qPCR data accurately reflect the measure you are interested in.
Many variables may negatively impact results: use of non-validated reference genes, non-specific assays, cDNA contaminated with trace amounts of genomic DNA that may be co-amplified, non-calibrated pipets or instruments with too large well-to-well variation. In this blog, I will describe an easy method to assess instrument related measurement errors.
Why functional validation beats temperature probes
Both the optics and the temperature control of a qPCR instrument can affect the outcome. In a quality controlled environment, it is common to have a cycler’s temperature being tested with calibrated temperature probes, either as a service or by investing in a monitoring system. This comes at a relative high cost and ignores the fact that other aspects may also affect the quality of the data being generated by a qPCR instrument.
That is why I here describe a functional validation test that is very affordable and can be easily run in every lab without the need for specific equipment or extensive training. We call this "run homogeneity" testing because it is based on the assessment of the homogeneity (or the lack thereof) of the results generated by your qPCR instrument.
How run homogeneity testing works
The principle of run homogeneity testing is quite straightforward. A single qPCR mix containing all the required elements for a qPCR reaction (mastermix, template, primers (and probe)) is made and distributed across an entire plate, chip, rotor, or array, collectively called "run" according to RDML nomenclature.
I advise to use standard conditions (same mix, same volume), an assay with demonstrated performance and an assay-sample combination that yields Cq values around 25. Very low Cq values are probably not representative and too high Cq values suffer from Poisson noise that may be incorrectly interpreted as suboptimal instrument performance. Such a test can be set up and analyzed with less than half a day of work and for a cost of less than €100.
I encourage to run this functional validation test when acquiring a new instrument, upon instrument maintenance, in case of concerns about instrument performance (e.g. issues observed) and on a regular basis (e.g. quarterly). The regular testing does not only serve as preventive measure but also allows to perform trend analysis; is there a gradual decline in instrument performance that may otherwise have remained undetected?
Analyzing the results
Analysis of a run homogeneity test should include at least an evaluation of Cq values, but may also include Tm values (for DNA binding dye assays) or end point fluorescence values. We typically combine an objective, statistical data analysis with a visual inspection of a plate heat map to screen for spatial effects.
Relevant statistics are the standard deviation on the Cq values (representing overall variability), the number of missing values (may be missed in variance analysis) and the number of data points outside of a given Cq interval around the median Cq.
Data can be processed in a spreadsheet, but the plate heat map may be more challenging for some users. To facilitate and standardize these analyses, Clarida offers run homogeneity testing as a free Community tool.
What to aim for
The interpretation of results may be relative (among instruments or in function of time (i.e. trend analysis)) or absolute (based on custom requirements). As always, universal absolute requirements cannot be given because they are determined by your instrument and reagent choice, the number of PCR replicates in a real study and, more importantly, by your experiment specific requirements (researching big differences or diagnosing small changes).
I would personally be quite strict on missing values (zero tolerance if repeated tests indicate it is an instrument issue and not a mere pipetting mistake), aim for a standard deviation below 0.2 or 0.25, and require 95% and 100% of replicates to fall within 0.5 and 1 cycle, respectively. The smaller Cq interval is used because it matches our ongoing PCR replicate quality control. The latter, higher interval because there should not be any replicates deviating to such a degree.
Key takeaways
- Run homogeneity testing is a low-cost, practical method to assess qPCR instrument performance - no specialized equipment required.
- Distribute a single qPCR mix across an entire plate and evaluate Cq consistency, spatial effects, and missing values.
- Aim for Cq values around 25, standard deviation below 0.25, and zero tolerance for missing values.
- Run the test when acquiring a new instrument, after maintenance, and quarterly for trend analysis.
- Clarida offers run homogeneity testing as a free Community tool at app.clarida.bio/plate-homogeneity.
Jan Hellemans
Co-founder, Clarida